ptical Powermeter by Ophir NOVA, indicating power 30.67 mW. [1]

Original filename: https://www.custom-cal.com/images/instruments/312/NOVA%20II.jpg

Description

OPHIR NOVA II Description / Specification:

OPHIR NOVA II Laser Power / Energy Meter

The Ophir Nova II is a microprocessor-based Laser Power/Energy Meter providing a broad range of measurements, displays, and data handling options. It operates with thermopile, pyroelectric and photodiode sensors, and uses smart connector technology.

Input Specifications Thermal, Photodiode. Input Ranges: 15 nA - 1.5 mA full scale in 16 ranges. A to D Sampling rate: 15Hz. A to D Resolution: 23 bits plus sign (0.00001%). A to D Linearity: better than 17 bits (0.0009%. Electrical accuracy: ±0.25% ± 20 pA new; ±0.5% ±50 pA after 1 year. Electrical input noise level: 500 nV or 1.5 pA + 0.0015% of input range @3Hz. Dynamic range: 9 decades (1:109).

Input Specifications Pyroelectric Sensors. Input Range: 0 - 6V full scale. A to D Sampling rate: 4 KHz. A to D resolution: 12 bits no sign (0.025% resolution). Electrical accuracy: ±0.25% new; ±0.5% after 1 year. Electrical input noise: 2 mV.

General Specifications. Detector Compatibility: Thermopile, photodiode and pyroelectric. Analog output 4 scales, user-configurable: 1v,2v,5v,10v full-scale; 0.03% resolution, 100 ohms impedance. Analog output accuracy: ±0.2% (of reading) ±0.3% of full scale volts.

References

  1. https://www.custom-cal.com/ModelInfo.aspx?kn=22849&m=OPHIR_NOVA%20II&srv=Calibration OPHIR NOVA II CALIBRATION and OPHIR NOVA II REPAIR (2021)
    A calibration by Custom-Cal is performed by engineers with extensive OEM experience. We have the expertise and the necessary standards to perform the OPHIR NOVA II Calibration, onsite calibration may be available. We specialize in quick turnaround times and we can handle expedited deliveries upon request. ..

File history

Click on a date/time to view the file as it appeared at that time.

Date/TimeThumbnailDimensionsUserComment
current16:57, 9 July 2021Thumbnail for version as of 16:57, 9 July 2021210 × 400 (15 KB)T (talk | contribs)ptical Powermeter by Ophir NOVA, indicating power 30.67 mW. <ref> https://www.custom-cal.com/ModelInfo.aspx?kn=22849&m=OPHIR_NOVA%20II&srv=Calibration OPHIR NOVA II CALIBRATION and OPHIR NOVA II REPAIR (2021)<br> A calibration by Custom-C...